User guide
To set the value of the calibration resistor, include the line
CALRES=resistance
Güralp Systems digitisers normally use a 51 kΩ resistor
(CALRES=51000).
To set the sensor type, include the line
TYPE=model-number
e.g. 3T, 5T, etc..
To set the response of the sensor, include the line
RESPONSE=response-type unit
Some of the values you can use are given in the table below. If
none of these match the response of your instrument, please
contact support for advice.
Sensor Sensor type code
Units
(V/A)
CMG-5T or 5TD,
DC – 100 Hz response
CMG-5_100HZ A
CMG-40T-1 or 6T-1,
1 s – 100 Hz response
CMG-40_1S_100HZ V
CMG-40T-1 or 6T-1,
2 s – 100 Hz response
CMG-40_2S_100HZ V
CMG-40T-1 or 6T-1,
10 s – 100 Hz response
CMG-40_10S_100HZ V
CMG-40T, 20 s – 50 Hz response
CMG-40_20S_50HZ V
CMG-40T, 30 s – 50 Hz response
CMG-40_30S_50HZ V
CMG-3T or 3ESP,
30 s – 50 Hz response
CMG-3_30S_50HZ V
CMG-40T, 60 s – 50 Hz response
CMG-40_60S_50HZ V
CMG-3T or 3ESP,
60 s – 50 Hz response
CMG-3_60S_50HZ V
CMG-3T or 3ESP,
100 s – 50 Hz response
CMG-3_100S_50HZ V
CMG-3T or 3ESP,
120 s – 50 Hz response
CMG-3_120S_50HZ V
January 2014 23